(Circulation. 1999;99:1516-1522.)
© 1999 American Heart Association, Inc.
Basic Science Reports |
From the University of Technology, Department of Cardiology (P.N.S., K.Z., M.W., F.B., K.M., P.H., C.S.) and Department of Cardiovascular Surgery (F.A.S., M.G.), Aachen, Germany.
Correspondence to Dr Patrick N. Schauerte, Cardiovascular Section, Department of Internal Medicine, University of Oklahoma Health Sciences Center, D.V.A. Medical Center/Research Service 151-F, 921 NE 13th St, Oklahoma City, OK 73104. E-mail Patrick-Schauerte{at}ouhsc.edu
| Abstract |
|---|
|
|
|---|
Methods and ResultsDFT and ULV were determined by up-down protocols for 10 BSDs. ULVs were measured by T-wave scanning during ventricular pacing (cycle length 500 ms). In protocol 1 (n=6), a pectoral "active can" was combined with an electrode in the superior vena cava as common cathode and a right ventricle electrode as anode (AC+SVC). In protocol 2 and protocol 3 (n=6 each), only the "active can" was used as proximal electrode (AC). Capacitance was 150 µF during both phases in protocol 1 and protocol 3 but 150 µF (phase 1) and 300 µF (phase 2) in protocol 2. ULV and DFT demonstrated a linear correlation in each protocol (r=0.78 to 0.84). Lowest DFTs were found at 10 ms for AC+SVC and at 14 ms for AC (P<0.001). At optimal BSDs, voltage DFTs did not differ significantly between AC (527±57 V) and AC+SVC (520±70 V). Switching capacitors for phase 2 in a way that reduced leading-edge voltage by 50% while doubling capacity did not change BSD for optimal voltage DFT but increased minimum DFT from 527±57 V to 653±133 V (P=0.04).
ConclusionsThe BSD with lowest DFT is shorter for AC+SVC than for AC. There is no significant difference in voltage DFT between both at optimal BSD. A lower phase 2 capacitance reduces DFTs irrespective of BSD. Because strength-duration curves for DFT and ULV correlate for different BSDs, lead systems, and phase 2 capacitances, ULV determination may allow the prediction of waveforms with lowest DFT.
Key Words: defibrillation fibrillation waves
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Electrode Configuration and Capacitors
Protocol 1
In 6 pigs, a 3-electrode system was applied. The RVA coil served
as anode during phase 1 of the biphasic shock and the SVC coil together
with the ICD shell as cathode (AC+SVC). Shocks were delivered by a
150-µF capacitor (two 300-µF capacitances connected in series;
Ventritex-HVS-02, Ventritex Corp) during both shock phases (phase
separation, 200 µs). Phase 2 leading-edge voltage equaled phase 1
trailing-edge voltage.
Protocol 2
In 6 pigs, shocks were delivered between the RVA electrode and
the pectoral ICD shell (AC). The RVA electrode was anode during phase
1. Shocks were delivered by a 150-µF capacitor during phase 1 (two
300-µF capacitances connected in series) and a single 300-µF
capacitor during phase 2 (phase separation, 200 µs;
Ventritex-HVS-02). Leading-edge voltage of phase 2 was half the
trailing-edge voltage of phase 1.
Protocol 3
In 6 pigs, shocks were delivered between the RVA electrode and
the pectoral ICD shell (AC) as in protocol 2, whereas the waveform was
identical to protocol 1.
Waveform Durations
DFTs and ULVs were determined for 10 TSDs (2, 4, 6, 8, 10, 12,
14, 16, 18, and 20 ms), with a relation of phase 1 to phase 2 duration
of 50%:50%. The order of testing of the different TSDs was randomized
within each animal. Peak voltage and current during phase 1 were
recorded by a digital oscilloscope. Pathway impedance was
determined by dividing peak voltage by peak current. The voltage
decrease during phase 1 (phase 1 tilt) was calculated according to the
formula phase 1
tilt=(1-V2/V1)x100 (%),
where V1=phase 1 leading-edge voltage and
V2=phase 1 trailing-edge voltage. The phase 1
trailing-edge voltage was estimated by use of the formula
Vt=V0xet/RxC,
where Vt=capacitor voltage as a function of time
t; V0=initial capacitor voltage, at t=0;
R=impedance; and C=capacitance.
ULV Measurement
ULV determinations were performed during rapid pacing through
the tip and ring of the RVA electrode. A train of 8 stimuli (cycle
length 500 ms) was delivered at twice the diastolic pacing
threshold. Six surface ECG limb leads were recorded at a paper
speed of 100 mm/s (Mingograph, Siemens Corp) to select the lead
with the latest peaking monophasic T-wave during pacing. The time from
stimulus to the peak of this T-wave was measured and averaged during 3
consecutive beats. Throughout the whole study, new measurements of the
stimulus to T-wave interval were taken every 5 VF episodes or before
ULV determination for a new TSD started. ULV testing started with a
shock strength of 710 V delivered to the peak of the T-wave of the 8th
paced ventricular beat. A step-upstep-down protocol with
a step-up size of 40 V, a step-down size of 80 V, and final steps of 20
V was used. The lowest voltage that did not induce VF was defined as
the ULV. After ULV determination for the peak of the T-wave, the same
protocol was started for shocks delivered at peak -20 ms, peak -40
ms, and peak +20 ms. Thus 4 ULVs were determined for each TSD in each
animal. The highest of the 4 ULVs was defined as
ULVscan for each shock duration and pig. One
minute elapsed between shocks that did not induce VF and subsequent
shocks.
DFT Measurement
The DFTs were determined during VF episodes induced while
testing the ULV for the same TSD. DFT testing started with a shock
strength of 710 V by using the same step-upstep-down protocol as for
the ULV determination. Defibrillating shocks were delivered 10 seconds
after the onset of VF. In case of failure, an internal 990-V shock was
applied. After each VF episode, a next shock was not delivered within 3
minutes and before blood pressure and heart rate had returned to
normal. In addition, the DFT was determined for shocks with a TSD of 10
ms at the beginning of the experiment and hourly thereafter. If the DFT
voltage differed by >40 V for this control waveform, the pig was
excluded from analysis.
Statistical Analysis
Data are expressed as mean±1 SD. Repeated-measures ANOVA was
used to compare peak voltage and current at the DFT or
ULVscan among the 10 waveforms in each protocol.
For comparison of DFTs/ULVscan between protocols,
a 2-tailed unpaired Student's t test was used. Individual
DFT and ULVscan values of all animals and TSDs in
each protocol were pooled for calculation of linear regression. In each
protocol, the DFT/ULVscan values for each TSD
were compared with the use of a paired t test. Differences
in impedances between protocols were assessed by 2-tailed unpaired
Student's t tests. The percentages of coupling intervals
relative to the peak of the T-wave that identified the
ULVscan were compared by use of a
2 test. ULV values for different coupling
intervals were compared with the use of a paired t test.
Probability values
0.05 were considered statistically
significant.
| Results |
|---|
|
|
|---|
|
Protocol 1
The shock strength-duration curve for the DFT demonstrated a
single minimum of peak voltage at a TSD of 10 ms (phase 1 tilt, 66%;
P<0.001, Figure 1
). The same
applied for the DFT in terms of peak current (Table
). Similarly, the
strength-duration curve for the ULVscan showed a
nadir at a TSD of 10 ms (phase 1 tilt, 64%; P<0.001,
Figure 1
). At each TSD, the ULVscan was
lower than the DFT (P<0.05 except for a TSD of 4 ms,
P=NS). A linear correlation existed between ULV and DFT,
with r=0.78 (Figure 4A
). The impedance did not differ
significantly between DFT shocks (31±3
) and
ULVscan shocks (30±3
).
|
|
Protocol 2
The lowest voltage DFT was observed at a TSD of 14 ms (phase 1
tilt, 75%; P<0.001, Figure 2
). Likewise, the lowest
ULVscan was found at a TSD of 14 ms (phase 1
tilt, 70%; P<0.001, Figure 2
). Basically the
same results were observed for the DFT and
ULVscan in terms of peak current (Table
). At each
TSD, the ULVscan was lower than the DFT
(P<0.05 except for a TSD of 6 and 20 ms, P=NS).
There was a linear correlation between ULV and DFT, with
r=0.84 (Figure 4B
). The defibrillation impedance was
37±8
compared with 38±7
for ULV shocks (P=NS).
Both impedances were higher compared with AC+SVC during protocol 1
(each P<0.001).
|
Protocol 3
The lowest voltage DFTs and ULVscan were
present at a TSD of 14 ms (P<0.001, Figure 3
). The corresponding phase 1 tilts were
71% for DFT shocks and 69% for ULV shocks. The DFT and
ULVscan values in terms of peak current were also
lowest at 14 ms. At each TSD, the ULVscan was
consistently lower than the DFT (P<0.05 except for
a TSD of 4 ms, P=NS). The linear regression line is shown in
Figure 4C
(r=0.8). The
impedance for DFT shocks was 36±5
compared with 38±7
during
ULVscan determination (P=NS).
|
Comparison Between Waveforms of Protocol 2 and Protocol 3
The only difference between protocol 2 and protocol 3 was the use
of a 300-µF capacitor during phase 2 of the biphasic pulse in
protocol 2, resulting in reducing the leading-edge voltage of phase 2
to half the value used in the other protocols as a result of the
technical method used (removing one of the two 300-µF capacitors
switched in series to obtain 150 µF for phase 1). The optimal TSD was
equal for both waveforms (14 ms). However, the voltage DFT was
significantly lower for the 14-ms waveform in protocol 3 (527±57 V)
compared with the 14-ms waveform in protocol 2 (653±133 V,
P=0.04). The same holds true for the peak current at the DFT
(protocol 3, 14±2 A vs protocol 2, 18±5 A; P=0.05). These
results would have been predicted by the ULVscan:
Both voltage ULVscan were lowest at a TSD of 14
ms. At this TSD, protocol 3 rendered a significantly lower
ULVscan (403±75 V) than protocol 2 (553±124 V,
P=0.02). Similar results were observed for the
ULVscan in terms of peak current (protocol 3 at
TSD of 14 ms, 11±2 A vs protocol 2 at TSD of 14 ms, 15±5 A;
P=0.04). A significant difference between both protocols was
not found for the defibrillation impedance or for the impedance of ULV
shocks.
Comparison Between Electrode Configurations of Protocol 1 and
Protocol 3
The difference between protocol 1 and protocol 3 was the
incorporation of an additional defibrillation electrode in the SVC
connected as common cathode with the pectoral ICD shell in protocol 1
(AC+SVC). This resulted in a significantly lower mean impedance for
both DFT and ULVscan for AC+SVC compared with the
AC system in protocol 3 (P<0.001). Moreover, the TSD with
lowest DFT was shorter for AC+SVC (10 ms) than for AC (14 ms). There
was an insignificant trend toward lower DFTs for AC+SVC as opposed to
AC for TSDs, being
10 ms (Figure 5
). By
contrast, except for a TSD of 20 ms, the DFTs for TSDs
12 ms were
higher for AC+SVC compared with AC. This reached statistical
significance at a TSD of 14 ms (P=0.05). Comparing both
electrode configurations each at the TSD with the lowest DFT, no
significant difference was present between both (AC+SVC, 10 ms:
520±70 V vs AC, 14 ms: 527±57 V, P=NS). Of interest,
because of a lower impedance of the AC+SVC configuration, peak current
for AC+SVC at 10 ms (18±2 A) was significantly higher than for AC at
14 ms (14±2 A, P=0.02). The results of
ULVscan determination paralleled those of the
DFT measurements: Whereas the lowest ULVscan was
observed for a TSD of 10 ms during protocol 1 (433±53 V), the
ULVscan was minimal for a TSD of 14 ms during
protocol 3 (403±75 V). Similar to the DFT measurements, differences
of peak voltage at the ULVscan between AC and
AC+SVC at optimal TSD were not significant. Also, peak current at the
ULVscan was significantly higher for AC+SVC at 10
ms (15±2 A) than for AC at 14 ms (11±2 A, P=0.002).
|
Coupling Intervals at the ULVscan
The mean stimulus to T-wave interval was 382±23 ms. Shocks
delivered to the peak of the T-wave were more likely to identify the
ULVscan than were shocks of other coupling
intervals (peak of T-wave: 48% vs peak-20 ms: 29%, peak+20 ms: 30%
or peak-40 ms: 3%, P<0.001 each, n=180). In those
instances in which a coupling interval of peak-40 ms identified the
ULVscan, the ULVs at peak-40 ms were higher
(594±138 V) than at a coupling interval of peak-20 ms (546±114 V,
P=NS). When a coupling interval of peak+20 ms identified the
ULVscan, the ULV at peak+20 ms (577±159 V) was
significantly higher than the ULVs at the peak of the T-wave (490±137
V, P<0.001).
| Discussion |
|---|
|
|
|---|
The DFT data of the present study may be important with respect to some recent studies on different pectoral lead systems in patients. Whereas one study17 did not find a significant DFT difference between AC and AC+SVC (60% phase 1 tilt, both), another study18 reported a decrease of the DFT for the AC+SVC system as compared with the AC system (65% phase 1 tilt, both). It is important to note that the pathway impedance was significantly lower for AC+SVC than for AC in both clinical studies17 18 and our animal study. In fact, theoretical concepts suggest that with decreasing capacitance and impedance, the shock duration with the lowest DFT decreases.15 19 As predicted, the TSD with minimal DFT voltage in our study was shorter for the lower impedance AC+SVC system (10 ms) than for the higher impedance AC configuration (14 ms). The respective phase 1 tilt was slightly higher for AC (71%) compared with AC+SVC (66%). The latter would not have been expected from a theoretical standpoint, which predicts that the phase 1 tilt with the lowest DFT increases with decreasing impedance and capacitance.15 Notably, for time constants T (T=impedancexcapacitance) between 2 and 12 ms, changes of the optimal phase 1 tilt were suggested to be less pronounced than changes of phase 1 duration.15 The phase 1 tilt in the current study was calculated by using the impedance measured at the leading edge of phase 1. This is only an estimate of the real phase 1 tilt as it would have been obtained by measuring phase 1 leading and trailing-edge voltage. In fact, the impedance has been reported to change during phase 1 of biphasic shocks.20 Therefore, one could imagine that the tilts calculated in the present study may not have exactly reflected the real voltage drop during phase 1. This may partly explain why the tilts found in the present study differed from the tilts that would have been predicted on the basis of mathematical models. The results of the present study indicate that the comparison of lead systems with different impedances by using the same phase 1 tilt or shock duration may in fact misestimate the DFT difference. As such, the use of the same phase 1 tilt within each study but different phase 1 tilts between the studies may have contributed to the diverging results in recent studies that investigated the DFTs for an AC+SVC system as compared with an AC system.17 18
Another intention of our study was to compare the DFT and the ULV for different capacitances during phase 2 of the biphasic shocks. Changing phase 2 capacitance from 90 to 22.5 µF (phase 1, 90 µF) has previously been shown to decrease the DFT in patients.21 However, besides different phase 2 capacitances and phase 2 leading-edge voltages, differences in phase 1 duration might have contributed to the lower DFT for the 22.5-µF second-phase waveform. The present study showed that the TSD and phase 1 duration with the lowest DFT was equal for waveforms with different phase 2 capacitances. At optimal shock duration, the DFT for the 150-µF second phase was significantly lower than for the 300-µF second phase. It is important to note that besides differences of phase 2 capacitances there also was a difference of phase 2 leading-edge voltage for the 2 waveforms. This renders it difficult to explain by which mechanism the waveform used in protocol 3 decreased the DFT compared with protocol 2. A model developed by Kroll15 proposed that phase 2 unloads the residual charge that is left on the cardiac cell membrane after delivery of the first phase ("charge-burping" hypothesis). This residual charge is thought to cause postshock arrhythmias, thus diminishing defibrillation success.15 For a fixed pulse duration, the charge delivered by a capacitor equals CxV0x[1-e-t/RxC] (C=capacitance; V0=initial capacitor voltage, at t=0; t=pulse duration; R=impedance). Comparing protocol 2 and protocol 3, the product CxV0 remains constant (halving C, doubling V). Thus for an equal pulse duration and a similar impedance, a lower C value means more charge delivered. Conceptually, this could mean that protocol 2, with the higher C, provides less effective "charge-burping," causing higher DFTs. However, we did not specifically test this hypothesis, which remains to be checked in the Kroll model in a quantitative way and then verified in experiments that vary capacity and leading-edge voltage for phase 2 independently.
Limitations
In our study, the ULVscan was significantly
lower than the DFT. In dogs,22 it was demonstrated that
with rapid ventricular pacing the ULV approximately equaled
the DFT, whereas it was lower than the DFT when the pacing cycle length
was increased. Therefore, the choice of a faster pacing rate in our
study might have resulted in a ULVscan level
closer to the DFT level. In addition, only limited T-wave scanning was
performed in our study, which may have caused an underestimation of the
ULVscan. Additional T-wave shocks delivered at
peak+40 ms or peak-50 ms might have resulted in higher
ULVscan values.
To determine most accurately the DFT and ULV strength-duration relation, we used a step upstep-down protocol with final steps of 20 V. This required a maximum of 3 VF inductions to determine the ULV for a single time point of the T-wave. Recent clinical studies7 8 reported that other ULV protocols need only 1 or 2 VF inductions for ULV determination. Moreover, during clinical ULV testing, larger voltage steps will require fewer VF inductions, thus providing higher safety during ULV and DFT testing. This may, however, worsen the correlation between the ULV and DFT.
In this animal model, the interval from the pacing stimulus to the peak of the latest peaking monophasic T-wave was 382±23 ms and therefore longer than reported in a clinical study (358±56 ms) with the same pacing rate.23 However, all patients in that study had organic heart disease, which may have accounted for this difference. The lack of organic heart disease also may have affected the level of the ULVscan and the correlation between ULVscan and DFT in the present study.
No pig had to be excluded from analysis because of unstable DFT for the control waveform. Therefore, the number of previous shocks did not affect subsequent DFT results for this control waveform. Because we did not repeatedly determine the ULV for a control waveform throughout the experiment, we do not know whether the same might have applied for the ULV results.
Conclusions
The TSD for lowest DFT is
40% longer for AC (36
) than for
AC+SVC (31
). The respective tilts are more similar for these 2
configurations: 71% for AC and 66% for AC+SVC. At optimal TSD, the
DFT did not differ significantly between both lead systems.
Application of a 150-µF capacitor during both phases of the biphasic shocks led to a significant DFT reduction compared with shocks with a 150-µF first phase and a 300-µF second phase. The latter waveform is currently applied in some clinical devices. A lower phase 2 capacitance did not affect the TSD with the lowest DFT.
The DFT results would have been predicted by ULVscan determinations because strength-duration curves for DFT and ULVscan correlated for different electrode configurations and phase 2 capacitances. If results in patients are shown to be similar to our results in pigs, ULV determination may allow comparison of different electrode systems and capacitances, each at optimal TSD without the need for extensive DFT testing.
Received June 16, 1998; revision received October 14, 1998; accepted November 18, 1998.
| References |
|---|
|
|
|---|
2.
Schauerte P, Schöndube FA, Grossmann M,
Dörge H, Stein F, Dohmen B, Moumen A, Erena K, Messmer BJ,
Hanrath P, Stellbrink C. Influence of phase duration of biphasic
waveforms on defibrillation energy requirements with a 70 µF
capacitance. Circulation. 1998;97:20732078.
3. Leonelli FM, Kroll MW, Brewer JE. Defibrillation thresholds are lower with smaller storage capacitors. Pacing Clin Electrophysiol. 1995;18:16611665.[Medline] [Order article via Infotrieve]
4.
Bardy GH, Poole JE, Kudenchuk PJ, Dolack GL, Mehra R,
DeGroot P, Raitt MH, Jones GK, Johnson G. A prospective randomized
comparison in humans of biphasic waveform 60-µF and 120-µF
capacitance pulses using a unipolar defibrillation system.
Circulation. 1995;91:9195.
5.
Swerdlow CD, Kass RM, Chen P-S, Hwang C, Raissi S.
Effect of capacitor size and pathway resistance on defibrillation
threshold for implantable defibrillators. Circulation. 1994;90:18401846.
6. Steinbeck G, Dorwarth U, Mattke S, Hoffmann E, Markewitz A, Kaulbach K, Tassani P. Hemodynamic deterioration during ICD implant: predictor of high risk patients. Am Heart J. 1994;127(pt 2):10641067.
7. Swerdlow CD, Kass RM, O'Connor ME, Chen P-S. Effect of shock waveform on relationship between upper limit of vulnerability and defibrillation. J Cardiovasc Electrophysiol. 1998;9:339349.[Medline] [Order article via Infotrieve]
8.
Swerdlow CD, Peter CT, Kass RM, Gang ES, Mandel
WJ, Hwang C, Martin DJ, Chen P-S. Programming of implantable
cardioverter-defibrillators on the basis of the upper limit of
vulnerability. Circulation. 1997;95:14971504.
9.
Souza JJ, Malkin RA, Ideker RE. Comparison of upper
limit of vulnerability and defibrillation probability of success curves
using a nonthoracotomy lead system. Circulation. 1995;91:12471252.
10.
Chen P-S, Feld GK, Kriett JM, Mower MM, Tarazi RY,
Fleck P, Swerdlow CD, Gang ES, Kass RM. Relation between upper limit of
vulnerability and defibrillation thresholds in humans.
Circulation. 1993;88:186192.
11. Fan W, Gotoh M, Chen P-S. Effects of the pacing site, procainamide and lead configuration on the relationship between the upper limit of vulnerability and the defibrillation threshold. Pacing Clin Electrophysiol. 1995;18:12791284.[Medline] [Order article via Infotrieve]
12.
Topham SL, Cha Y-M, Peters BB, Chen P-S. Effects of
lidocaine on relation between defibrillation threshold and upper limit
of vulnerability in open-chest dogs. Circulation. 1992;85:11461151.
13. Wiggers CJ, Wegria R. Ventricular fibrillation due to single, localized induction and condenser shocks applied during the vulnerable phase of ventricular systole. Am J Physiol. 1940;128:500505.
14.
Chen P-S, Shibata N, Dixon EG, Martin RO, Ideker RE.
Comparison of defibrillation threshold and the upper limit of
ventricular vulnerability. Circulation. 1986;73:10221028.
15. Kroll MW. A minimal model of the single capacitor biphasic defibrillation waveform. Pacing Clin Electrophysiol. 1994;17:1:17821792.[Medline] [Order article via Infotrieve]
16.
Huang J, KenKnight BH, Walcott GP, Rollins DL, Smith
WM, Ideker RE. Effects of transvenous electrode polarity and waveform
duration on the relationship between defibrillation threshold and upper
limit of vulnerability. Circulation. 1997;96:13511359.
17.
Bardy GH, Dolack GL, Kudenchuk PJ, Poole JE, Mehra R,
Johnson G. Randomized comparison in humans of a unipolar defibrillation
system with that using an additional superior vena cava electrode.
Circulation. 1994;89:10901093.
18.
Gold MR, Olsovsky MR, Pelini MA, Peters RW, Shorofsky
SR. Comparison of single- and dual-coil active pectoral defibrillation
lead systems. J Am Coll Cardiol. 1998;31:13911394.
19. Cleland BG. A conceptual basis for defibrillation waveforms. Pacing Clin Electrophysiol. 1996;19:11861195.[Medline] [Order article via Infotrieve]
20. Tang ASL, Yabe S, Wharton JM, Dolker M, Smith WM, Ideker RE. Ventricular defibrillation using biphasic waveforms: the importance of phasic duration. J Am Coll Cardiol. 1989;13:207214.[Abstract]
21.
Yamanouchi Y, Mowrey KA, Nadzam GR, Hills DG, Kroll MW,
Brewer JE, Donohoo AM, Wilkoff BL, Tchou PJ. Large change in voltage at
phase reversal improve biphasic defibrillation thresholds.
Circulation. 1996;94:17681773.
22.
Malkin RA, Idriss SF, Walker RG, Ideker RE. Effect of
rapid pacing and T-wave scanning on the relation between the
defibrillation and upper-limit-of-vulnerability dose-response curves.
Circulation. 1995;92:12911299.