(Circulation. 1999;100:826-831.)
© 1999 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Department of Cardiology, Cleveland Clinic Foundation, Cleveland, Ohio, and SurVivaLink Corp, Minneapolis, Minn (J.E.B., K.F.O.).
Correspondence to Patrick J. Tchou, MD, Director, Clinical Cardiac Electrophysiology, Department of Cardiology/F15, Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195. E-mail tchoup{at}cesmtp.ccf.org
BackgroundPhase-2 voltage and maximum pulse width are dependent on phase-1 pulse characteristics in a single-capacitor biphasic waveform. The use of 2 separate output capacitors avoids these limitations and may allow waveforms with lower defibrillation thresholds. A previous report also suggested that the optimal tilt may be >70%. This study was designed to determine an optimal biphasic waveform by use of a combination of 2 separate and fully (95% tilt) discharging capacitors.
Methods and ResultsWe performed 2 external defibrillation studies in a pig ventricular fibrillation model. In group 1, 9 waveforms from a combination of 3 phase-1 capacitor values (30, 60, and 120 µF) and 3 phase-2 capacitor values (0=monophasic, 1/3, and 1.0 times the phase-1 capacitor) were tested. Biphasic waveforms with phase-2 capacitors of 1/3 times that of phase 1 provided the highest defibrillation efficacy (stored energy and voltage) compared with corresponding monophasic and biphasic waveforms with the same capacitors in both phases except for waveforms with a 30-µF phase-1 capacitor. In group 2, 10 biphasic waveforms from a combination of 2 phase-1 capacitor values (30 and 60 µF) and 5 phase-2 capacitor values (10, 20, 30, 40, and 50 µF) were tested. In this range, phase-2 capacitor size was more critical for the 30-µF phase-1 than for the 60-µF phase-1 capacitor. The optimal combinations of fully discharging capacitors for defibrillation were 60/20 and 60/30 µF.
ConclusionsPhase-2 capacitor size plays an important role in reducing defibrillation energy in biphasic waveforms when 2 separate and fully discharging capacitors are used.
Key Words: defibrillation death, sudden ventricles