Circulation, Vol 82, 1467-1475, Copyright © 1990 by American Heart Association
H Baumgartner, S Khan, M DeRobertis, L Czer and G Maurer
To evaluate possible causes of discrepancy between Doppler and catheter
gradients across prosthetic valves, five sizes (19-27 mm) of St. Jude and
Hancock valves were studied in an aortic pulsatile flow model. Catheter
gradients at multiple sites distal to the valve were compared with
simultaneously obtained Doppler gradients. In the St. Jude valve,
significant differences between Doppler and catheter gradients measured 30
mm downstream from the valve were found: Doppler gradients exceeded peak
catheter gradients of 10 mm Hg or more by 81 +/- 35% (15 +/- 3.6 mm Hg),
and mean catheter gradients by 71 +/- 11% (10.3 +/- 2.5 mm Hg). When the
catheter was pulled back through the tunnel-like central orifice of the
valve, high localized gradients at the valve plane and significant early
pressure recovery were found. When the catheter was pulled back through the
large side orifices, gradients at the same level were only 46 +/- 6% of the
central orifice gradients (mean difference, 7.6 +/- 4.5 mm Hg). Doppler
peak and mean gradients showed excellent agreement with the highest central
orifice catheter gradients (mean difference, 1.0 +/- 3.1 and 0.9 +/- 1.5 mm
Hg, respectively). A significantly better agreement between Doppler and
catheter gradients at 30 mm was found for the Hancock valve, although
Doppler peak and mean gradients were still slightly greater than catheter
gradients. Doppler gradients exceeded catheter gradients by 18 +/- 10% (3.4
+/- 1.9 mm Hg) and 13 +/- 11% (2.1 +/- 0.9 mm Hg), respectively. When the
catheter was pulled back through the valve, the highest gradients were
found approximately 20 mm distal to the valve ring.(ABSTRACT TRUNCATED AT
250 WORDS)
ARTICLES
Discrepancies between Doppler and catheter gradients in aortic prosthetic valves in vitro. A manifestation of localized gradients and pressure recovery
Division of Cardiology and Cardiovascular Surgery, Cedars-Sinai Medical Center, Los Angeles, CA 90048.
This article has been cited by other articles:
![]() |
P. Pibarot and J. G. Dumesnil Prosthetic Heart Valves: Selection of the Optimal Prosthesis and Long-Term Management Circulation, February 24, 2009; 119(7): 1034 - 1048. [Full Text] [PDF] |
||||
![]() |
R. Guenzinger, W. B. Eichinger, I. Hettich, S. Bleiziffer, D. Ruzicka, R. Bauernschmitt, and R. Lange A prospective randomized comparison of the Medtronic Advantage Supra and St Jude Medical Regent mechanical heart valves in the aortic position: is there an additional benefit of supra-annular valve positioning? J. Thorac. Cardiovasc. Surg., August 1, 2008; 136(2): 462 - 471. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Enriquez-Sarano, V. T. Nkomo, and H. Michelena Principles and Practice of Echocardiography in Cardiac Surgery Card. Surg. Adult, January 1, 2008; 3(2008): 315 - 348. [Full Text] |
||||
![]() |
N. J. Howell, B. E. Keogh, V. Barnet, R. S. Bonser, T. R. Graham, S. J. Rooney, I. C. Wilson, and D. Pagano Patient-prosthesis mismatch does not affect survival following aortic valve replacement. Eur. J. Cardiothorac. Surg., July 1, 2006; 30(1): 10 - 14. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. M. Popescu, E. Prokop, J. A. Elefteriades, K. Kett, and P. G. Barash Phantom Aortic Valve Pressure Gradient: Discrepancies Between Cardiac Catheterization and Doppler Echocardiography Anesth. Analg., May 1, 2005; 100(5): 1259 - 1262. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. E. Richards, D. Deserranno, E. Donal, N. L. Greenberg, J. D. Thomas, and M. J. Garcia Influence of structural geometry on the severity of bicuspid aortic stenosis Am J Physiol Heart Circ Physiol, September 1, 2004; 287(3): H1410 - H1416. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Albes, M. Hartrumpf, V. Rudolph, T. Krempl, E. Huttemann, R. Vollandt, and T. Wahlers Are mechanical valves with enhanced inner diameter advantageous in the small sized aortic annulus? Ann. Thorac. Surg., November 1, 2003; 76(5): 1564 - 1570. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Vannan and K. Sarkar Assessment of mechanical aortic valve prosthesis by means of Doppler echocardiography: what to measure and why? J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 317 - 320. [Full Text] [PDF] |
||||
![]() |
R. A. Levine and E. Schwammenthal Stenosis is in the eye of the observer: impact of pressure recovery on assessing aortic valve area J. Am. Coll. Cardiol., February 5, 2003; 41(3): 443 - 445. [Full Text] [PDF] |
||||
![]() |
O. Bech-Hanssen, K. Caidahl, I. Wallentin, P. Ask, and B. Wranne Assessment of effective orifice area of prosthetic aortic valves with Doppler echocardiography: An in vivo and in vitro study J. Thorac. Cardiovasc. Surg., August 1, 2001; 122(2): 287 - 295. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Girard, F. A. Miller Jr., T. A. Orszulak, C. J. Mullany, S. Montgomery, W. D. Edwards, H. D. Tazelaar, J. F. Malouf, and A. J. Tajik Reoperation for prosthetic aortic valve obstruction in the era of echocardiography: trends in diagnostic testing and comparison with surgical findings J. Am. Coll. Cardiol., February 1, 2001; 37(2): 579 - 584. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pibarot and J. G. Dumesnil Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1131 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Becassis, M. Hayot, J.-M. Frapier, F. Leclercq, L. Beck, J.e. Brunet, E. Arnaud, C. Prefaut, P.-A. Chaptal, J. M. Davy, et al. Postoperative exercise tolerance after aortic valve replacement by small-size prosthesis: Functional consequence of small-size aortic prosthesis J. Am. Coll. Cardiol., September 1, 2000; 36(3): 871 - 877. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Garcia, P. Pibarot, J. G. Dumesnil, F. Sakr, and L.-G. Durand Assessment of Aortic Valve Stenosis Severity : A New Index Based on the Energy Loss Concept Circulation, February 22, 2000; 101(7): 765 - 771. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Bech-Hanssen, K. Caidahl, B. Wall, P. Myken, S. Larsson, and I. Wallentin INFLUENCE OF AORTIC VALVE REPLACEMENT, PROSTHESIS TYPE, AND SIZEON FUNCTIONAL OUTCOME AND VENTRICULAR MASS IN PATIENTS WITH AORTIC STENOSIS J. Thorac. Cardiovasc. Surg., July 1, 1999; 118(1): 57 - 65. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Baumgartner, T. Stefenelli, J. Niederberger, H. Schima, and G. Maurer "overestimation" of catheter gradients by doppler ultrasound in patients with aortic stenosis: a predictable manifestation of pressure recovery J. Am. Coll. Cardiol., May 1, 1999; 33(6): 1655 - 1661. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Y. Leung, J. Wong, L. Rodriguez, M. Pu, P. M. Vandervoort, and J. D. Thomas Application of Color Doppler Flow Mapping to Calculate Orifice Area of St Jude Mitral Valve Circulation, September 22, 1998; 98(12): 1205 - 1211. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. H. Marcus, R. S. Heinrich, J. Bednarz, S. Lupovitch, J. Abruzzo, R. Borok, B. Vandenberg, R. E. Kerber, W. Piccione, A. P. Yoganathan, et al. Assessment of Small-Diameter Aortic Mechanical Prostheses : Physiological Relevance of the Doppler Gradient, Utility of Flow Augmentation, and Limitations of Orifice Area Estimation Circulation, September 1, 1998; 98(9): 866 - 872. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Niederberger, H. Schima, G. Maurer, and H. Baumgartner Importance of Pressure Recovery for the Assessment of Aortic Stenosis by Doppler Ultrasound: Role of Aortic Size, Aortic Valve Area, and Direction of the Stenotic Jet In Vitro Circulation, October 15, 1996; 94(8): 1934 - 1940. [Abstract] [Full Text] |
||||
![]() |
J. R. Gonzalez-Juanatey, J. M. Garca-Acuna, M. V. Fernandez, A. A. Cendon, V. C. Fuentes, J. B. Garcia-Bengoechea, and M. G. de la Pena INFLUENCE OF THE SIZE OF AORTIC VALVE PROSTHESES ON HEMODYNAMICS AND CHANGE IN LEFT VENTRICULAR MASS: IMPLICATIONS FOR THE SURGICAL MANAGEMENT OF AORTIC STENOSIS J. Thorac. Cardiovasc. Surg., August 1, 1996; 112(2): 273 - 280. [Abstract] [Full Text] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini COMPARISON OF HEMODYNAMIC PERFORMANCES OF ST. JUDE MEDICAL AND CARBOMEDICS 21 MM AORTIC PROSTHESES BY MEANS OF DOBUTAMINE STRESS ECHOCARDIOGRAPHY J. Thorac. Cardiovasc. Surg., February 1, 1996; 111(2): 408 - 415. [Abstract] [Full Text] |
||||
![]() |
A. Laske, R. Jenni, M. Maloigne, G. Vassalli, O. Bertel, and M. I. Turina Pressure Gradients Across Bileaflet Aortic Valves by Direct Measurement and Echocardiography Ann. Thorac. Surg., January 1, 1996; 61(1): 48 - 57. [Abstract] [Full Text] |
||||
![]() |
P. M. Vandervoort, N. L. Greenberg, M. Pu, K. A. Powell, D. M. Cosgrove, and J. D. Thomas Pressure Recovery in Bileaflet Heart Valve Prostheses : Localized High Velocities and Gradients in Central and Side Orifices With Implications for Doppler-Catheter Gradient Relation in Aortic and Mitral Position Circulation, December 15, 1995; 92(12): 3464 - 3472. [Abstract] [Full Text] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini Evaluation of the Hemodynamic Performance of Small CarboMedics Aortic Prostheses Using Dobutamine-Stress Doppler Echocardiography Ann. Thorac. Surg., October 1, 1995; 60(4): 1048 - 1052. [Abstract] [Full Text] |
||||
![]() |
S. Khan, A. Chaux, J. Matloff, C. Blanche, M. DeRobertis, R. Kass, T. P. Tsai, A. Trento, S. Nessim, R. Gray, et al. The St. Jude Medical valveExperience with 1000 cases J. Thorac. Cardiovasc. Surg., December 1, 1994; 108(6): 1010 - 1020. [Abstract] [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1990 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |