(Circulation. 1996;94:1607-1612.)
© 1996 American Heart Association, Inc.
Articles |
the Section of Cardiovascular Medicine, Lahey-Hitchcock Medical Center, Burlington, Mass; the Department of Cardiology, New England Deaconess Hospital, Boston, Mass; and the Clinical Department, Cardiac Pacemakers Inc/Guidant, St. Paul, Minn.
Correspondence to Ferdinand J. Venditti, MD, Chief, Cardiovascular Medicine, Lahey-Hitchcock Medical Center, Burlington, MA 01805.
Background Reports have demonstrated a circadian variation in the incidence of acute myocardial infarction, ventricular arrhythmias, and sudden cardiac death. We tested the hypothesis that a similar circadian variation exists for defibrillation energy requirements in humans.
Methods and Results We reviewed the time of defibrillation threshold (DFT) measurements in 134 patients with implantable cardioverter-defibrillators (ICDs) who underwent 345 DFT measurements. The DFT was determined in 130 patients at implantation, in 121 at a 2 months, and in 94 at 6 months. All patients had nonthoracotomy systems. The morning DFT (8 AM to 12 noon) was 15.1±1.2 J compared with 13.1±0.9 J in the midafternoon (12 noon to 4 PM) and 13.0±0.7 J in the late afternoon (4 to 8 PM), P<.02. In a separate group of 930 patients implanted with an ICD system with date and time stamps for each therapy, we reviewed 1238 episodes of ventricular tachyarrhythmias treated with shock therapy. To corroborate the hypothesis that energy requirements for arrhythmia termination vary during the course of the day, we plotted the failed first shock frequency for all episodes per hour. There was a significant peak in failed first shocks in the morning compared with other time intervals (P=.02).
Conclusions There is a morning peak in DFT and a corresponding morning peak in failed first shock frequency. This morning peak resembles the peaks seen in other cardiac events, specifically sudden cardiac death. These findings have important implications for appropriate ICD function, particularly in patients with marginal DFTs.
Key Words: circadian rhythm defibrillation tachyarrhythmias
This article has been cited by other articles:
![]() |
A. R. J. Mitchell, P. A. R. Spurrell, K. Kamalvand, M. Higson, R. Shanmuganathan, N. R. Patel, and N. Sulke What is the optimal electrode configuration for atrial defibrillators in man? Europace, January 1, 2002; 4(1): 41 - 44. [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow Implantation of Cardioverter Defibrillators Without Induction of Ventricular Fibrillation Circulation, May 1, 2001; 103(17): 2159 - 2164. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. P. Zipes Warning: The Short Days of Winter May Be Hazardous to Your Health Circulation, October 12, 1999; 100(15): 1590 - 1592. [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |