(Circulation. 1996;94:1768-1773.)
© 1996 American Heart Association, Inc.
Articles |
the Department of Cardiology, Cleveland (Ohio) Clinic Foundation (Y.Y., K.A.M., G.R.N., D.G.H., B.L.W., P.J.T.), and Angeion Corporation (M.W.K., J.E.B., A.M.D.), Plymouth, Minn.
Correspondence to Patrick J. Tchou, MD, Director, Clinical Cardiac Electrophysiology, Department of Cardiology/F15, Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195. E-mail TCHOUP@cesmtp.ccf.ORG.
Background Multiple factors contribute to an improved defibrillation threshold of biphasic shocks. The leading-edge voltage of the second phase may be an important factor in reducing the defibrillation threshold.
Methods and Results We tested two experimental biphasic waveforms with large voltage changes at phase reversal. The phase 2 leading-edge voltage was twice the phase 1 trailing-edge voltage. This large voltage change was achieved by switching two capacitors from parallel to series mode at phase reversal. Two capacitors were tested (60/15 microfarads [µF] and 90/22.5 µF) and compared with two control biphasic waveforms for which the phase 1 trailing-edge voltage equaled the phase 2 leading-edge voltage. The control waveforms were incorporated into clinical (135/135 µF) or investigational devices (90/90 µF). Defibrillation threshold parameters were evaluated in eight anesthetized pigs by use of a nonthoracotomy transvenous lead to a can electrode system. The stored energy at the defibrillation threshold (in joules) was 8.2±1.5 for 60/15 µF (P<.01 versus 135/135 µF and 90/90 µF), 8.8±2.4 for 90/22.5 µF (P<.01 versus 135/135 µF and 90/90 µF), 12.5±3.4 for 135/135 µF, and 12.6±2.6 for 90/90 µF.
Conclusions The biphasic waveform with large voltage changes at phase reversal caused by parallel-series mode switching appeared to improve the ventricular defibrillation threshold in a pig model compared with a currently available biphasic waveform. The 60/15-µF capacitor performed as well as the 90/22.5-µF capacitor in the experimental waveform. Thus, smaller capacitors may allow reduction in device size without sacrificing defibrillation threshold energy requirements.
Key Words: defibrillation ventricles electrical stimulation
This article has been cited by other articles:
![]() |
K. A. Mowrey, Y. Cheng, P. J. Tchou, and I. R. Efimov Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform Europace, January 1, 2002; 4(1): 27 - 39. [Full Text] [PDF] |
||||
![]() |
Y. Yamanouchi, J. E. Brewer, K. F. Olson, K. A. Mowrey, T. N. Mazgalev, B. L. Wilkoff, and P. J. Tchou Fully Discharging Phases : A New Approach to Biphasic Waveforms for External Defibrillation Circulation, August 24, 1999; 100(8): 826 - 831. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. N. Schauerte, K. Ziegert, M. Waldmann, F. A. Schondube, F. Birkenhauer, K. Mischke, M. Grossmann, P. Hanrath, and C. Stellbrink Effect of Biphasic Shock Duration on Defibrillation Threshold With Different Electrode Configurations and Phase 2 Capacitances : Prediction by Upper-Limit-of-Vulnerability Determination Circulation, March 23, 1999; 99(11): 1516 - 1522. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Yamanouchi, J. E. Brewer, K. A. Mowrey, A. M. Donohoo, B. L. Wilkoff, and P. J. Tchou Optimal Small-Capacitor Biphasic Waveform for External Defibrillation : Influence of Phase-1 Tilt and Phase-2 Voltage Circulation, December 1, 1998; 98(22): 2487 - 2493. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Zhou, W. M. Smith, R. K. Justice, J. L. Wayland, and R. E. Ideker Transmembrane potential changes caused by monophasic and biphasic shocks Am J Physiol Heart Circ Physiol, November 1, 1998; 275(5): H1798 - H1807. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Schauerte, F. A. Schondube, M. Grossmann, H. Dorge, F. Stein, B. Dohmen, A. Moumen, K. Erena, B. J. Messmer, P. Hanrath, et al. Influence of Phase Duration of Biphasic Waveforms on Defibrillation Energy Requirements With a 70-µF Capacitance Circulation, May 26, 1998; 97(20): 2073 - 2078. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Yamanouchi, K. A. Mowrey, M. J. Niebauer, P. J. Tchou, and B. L. Wilkoff Additional Lead Improves Defibrillation Efficacy With an Abdominal `Hot Can' Electrode System Circulation, December 16, 1997; 96(12): 4400 - 4407. [Abstract] [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |