(Circulation. 1997;95:449-454.)
© 1997 American Heart Association, Inc.
Articles |
the Department of Internal Medicine (Cardiology Division) of the Washington (DC) Hospital Center; the Department of Pathology, University of Texas at San Antonio (F.O.T.); and the Department of Surgery, Georgetown University Medical Center, Washington, DC (M.F., S.S.C.).
Background In-stent restenosis results primarily from neointimal hyperplasia. This study evaluated the efficacy and the optimal mode of administration of angiopeptin, a somatostatin analogue with antiproliferative activity, in a porcine coronary in-stent restenosis model.
Methods and Results Forty pigs were randomly assigned to one of four groups (n=10 per group): (1) controls receiving saline infusion at the site of stent implantation via a local delivery catheter, (2) local treatment group receiving one-time treatment (200 µg angiopeptin) at the site of stent placement, (3) systemic treatment group receiving continuous angiopeptin over a 1-week period via a subcutaneous osmotic pump (200 µg/kg total dose), and (4) combined local and systemic treatment group. Then, one oversized Palmaz-Schatz stent (mean ratio of stent to artery diameters, 1.3:1) was implanted in the left anterior descending coronary artery. The degree of neointimal reaction was evaluated 4 weeks later by angiography (maximal percent diameter stenosis), intravascular ultrasound (total in-stent neointimal volume), and histology (maximal area stenosis). Systemic treatment produced the least neointimal hyperplasia and significantly reduced in-stent restenosis compared with the control group by all end points, despite similar degrees of injury. Angiography showed 25±17% versus 50±17% diameter stenosis in the systemic angiopeptin group versus the control group (P<.0001), intravascular ultrasound revealed 23±10 versus 58±27 mm3 neointimal volume in the systemic angiopeptin versus control group (P=.0002), and histology showed 41±16% versus 69±18% area stenosis (P=.0016) in the systemic angiopeptin versus control group. Plasma angiopeptin levels revealed rapid clearance (within 6 hours) after local therapy, whereas the levels persisted for up to 2 weeks in the systemic group.
Conclusions This study shows that continuous subcutaneous treatment with angiopeptin after stent implantation significantly reduces in-stent restenosis by inhibiting neointimal hyperplasia.
Key Words: stents restenosis angioplasty
This article has been cited by other articles:
![]() |
R. Hoffmann and G.S. Mintz Coronary in-stent restenosis--predictors, treatment and prevention Eur. Heart J., November 1, 2000; 21(21): 1739 - 1749. [PDF] |
||||
![]() |
T. Takagi, K. Yoshida, T. Akasaka, S. Kaji, T. Kawamoto, Y. Honda, A. Yamamuro, T. Hozumi, and S. Morioka Hyperinsulinemia during oral glucose tolerance test is associated with increased neointimal tissue proliferation after coronary stent implantation in nondiabetic patients: A serial intravascular ultrasound study J. Am. Coll. Cardiol., September 1, 2000; 36(3): 731 - 738. [Abstract] [Full Text] [PDF] |
||||
![]() |
J Gunn and D Cumberland Stent coatings and local drug delivery. State of the art Eur. Heart J., December 1, 1999; 20(23): 1693 - 1700. [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |