Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 1999;99:2815-2818

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Trumble, D. R.
Right arrow Articles by Magovern, J. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Trumble, D. R.
Right arrow Articles by Magovern, J. A.
Related Collections
Right arrow Arrythmias-basic studies
Right arrow Pulmonary biology and circulation

(Circulation. 1999;99:2815-2818.)
© 1999 American Heart Association, Inc.


Basic Science Reports

Copulsation Balloon for Right Ventricular Assistance

Preliminary Trials

Dennis R. Trumble, MS; Chong S. Park, MD; James A. Magovern, MD

From the Department of Cardiothoracic Surgery, Allegheny General Hospital, and Cardiothoracic Surgery Research, Cardiovascular and Pulmonary Research Institute, Pittsburgh, Pa.

Correspondence to Dennis R. Trumble, MS, Allegheny General Hospital, Cardiothoracic Surgery Research, 9th Floor, South Tower, 320 E North Ave, Pittsburgh, PA 15212. E-mail trumble{at}pgh.auhs.edu

Background—Options for management of acute right ventricular (RV) failure are limited. This report describes preliminary testing of a temporary RV assist device that acts by direct compression of the RV. The system comprises a pancake-shaped silicone balloon (5 cm diameter) connected to a drive console that delivers a 65-mL pneumatic pulse during cardiac systole.

Methods and Results—Initial in vivo tests were performed on 6 pigs (weight, 41±4 kg). RV wall motion and stroke volume were monitored via transesophageal echocardiography. Acute RV failure was created by graded right coronary ligation, which yielded a 63% reduction in RV stroke volume (39.9±8.2 to 14.7±1.9 mL; P<0.002). We secured the balloon over the RV free wall by attaching it to the edges of the opened pericardium. The sternum was then reapproximated, and data were collected with the device off and on (every beat). Device placement had no deleterious effect on RV function. Balloon activation returned RV stroke volumes to normal (37.8±9.2 mL) and increased mean pulmonary artery pressures from 13±2 to 16±3 mm Hg (P<0.01). RV compression did not induce or exacerbate tricuspid regurgitation. Mean aortic pressure improved from postinfarction levels but did not return to normal.

Conclusions—We conclude that the pulmonary circulation can be supported in the short term via cardiac compression and that balloon copulsation techniques for short-term RV failure should be tested in long-term models.


Key Words: heart-assist device • hemodynamics • balloon • heart failure • ventricles