From the Departments of Medicine/Cardiology (S.N., A.K., M.M.Z., G.R.M., R.K., A.R., T.L.D., H.A., H.C., R.D.B., D.A.B., A.C.L., H.R.H.), Radiology (D.A.B., H.R.H.), Radiation Oncology (R.C.S.), and Biomedical Engineering (R.D.B., A.C.L., H.R.H.), Johns Hopkins Hospital, Baltimore, Md. * To whom correspondence should be addressed. E-mail: snazarian{at}jhmi.edu.
Background—Compared with fluoroscopy, the current imaging standard of care for guidance of electrophysiology procedures, magnetic resonance imaging (MRI) provides improved soft-tissue resolution and eliminates radiation exposure. However, because of inherent magnetic forces and electromagnetic interference, the MRI environment poses challenges for electrophysiology procedures. In this study, we sought to test the feasibility of performing electrophysiology studies with real-time MRI guidance. Methods and Results—An MRI-compatible electrophysiology system was developed. Catheters were targeted to the right atrium, His bundle, and right ventricle of 10 mongrel dogs (23 to 32 kg) via a 1.5-T MRI system using rapidly acquired fast gradient-echo images ( Conclusions—To the best of our knowledge, this is the first study to report the feasibility of real-time MRI-guided electrophysiology procedures. This technique may eliminate patient and staff radiation exposure and improve real-time soft tissue resolution for procedural guidance.
Accepted on April 25, 2008
Feasibility of Real-Time Magnetic Resonance Imaging for Catheter Guidance in Electrophysiology Studies
Saman Nazarian MD*,
5 frames per second). Catheters were successfully positioned at the right atrial, His bundle, and right ventricular target sites of all animals. Comprehensive electrophysiology studies with recording of intracardiac electrograms and atrial and ventricular pacing were performed. Postprocedural pathological evaluation revealed no evidence of thermal injury to the myocardium. After proof of safety in animal studies, limited real-time MRI-guided catheter mapping studies were performed in 2 patients. Adequate target catheter localization was confirmed via recording of intracardiac electrograms in both patients.
Related Article:
Circulation 2008 118: 211-212.
This article has been cited by other articles:
![]() |
B. A. Hoffmann, A. Koops, T. Rostock, K. Mullerleile, D. Steven, R. Karst, M. U. Steinke, I. Drewitz, G. Lund, S. Koops, et al. Interactive real-time mapping and catheter ablation of the cavotricuspid isthmus guided by magnetic resonance imaging in a porcine model Eur. Heart J., November 5, 2009; (2009) ehp460v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. E. Saikus and R. J. Lederman Interventional Cardiovascular Magnetic Resonance Imaging: A New Opportunity for Image-Guided Interventions J. Am. Coll. Cardiol. Img., November 1, 2009; 2(11): 1321 - 1331. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Kramer and J. Narula Interventional CMR: Great Promise, but a Long Road Ahead J. Am. Coll. Cardiol. Img., November 1, 2009; 2(11): 1337 - 1338. [Full Text] [PDF] |
||||
![]() |
S. S. Kim, Z. M. Hijazi, R. M. Lang, and B. P. Knight The use of intracardiac echocardiography and other intracardiac imaging tools to guide noncoronary cardiac interventions. J. Am. Coll. Cardiol., June 9, 2009; 53(23): 2117 - 2128. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Nordbeck, W. R. Bauer, F. Fidler, M. Warmuth, K.-H. Hiller, M. Nahrendorf, M. Maxfield, S. Wurtz, W. Geistert, J. Broscheit, et al. Feasibility of Real-Time MRI With a Novel Carbon Catheter for Interventional Electrophysiology Circ Arrhythm Electrophysiol, June 1, 2009; 2(3): 258 - 267. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Akoum and N. F. Marrouche Real-time imaging in electrophysiology: from intra-cardiac echo to real-time magnetic resonance imaging Europace, May 1, 2009; 11(5): 539 - 540. [Full Text] [PDF] |
||||
![]() |
L. F. Tops and J. J. Bax The Year in Imaging Related to Electrophysiology J. Am. Coll. Cardiol. Img., April 1, 2009; 2(4): 498 - 510. [Full Text] [PDF] |
||||
![]() |
A. Pandit and N. F. Marrouche Cardiac magnetic resonance in the world of the cardiac electrophysiologist the road to real-time cardiac magnetic resonance. J. Am. Coll. Cardiol. Img., March 1, 2009; 2(3): 317 - 318. [Full Text] [PDF] |
||||
![]() |
S. Nazarian, D. A. Bluemke, and H. R. Halperin Applications of Cardiac Magnetic Resonance in Electrophysiology Circ Arrhythm Electrophysiol, February 1, 2009; 2(1): 63 - 71. [Full Text] [PDF] |
||||
![]() |
P. Kirchhof, M. Ozgun, S. Zellerhoff, G. Monnig, L. Eckardt, K. Wasmer, W. Heindel, G. Breithardt, and D. Maintz Diastolic isthmus length and 'vertical' isthmus angulation identify patients with difficult catheter ablation of typical atrial flutter: a pre-procedural MRI study Europace, January 1, 2009; 11(1): 42 - 47. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Knackstedt, P. Schauerte, and P. Kirchhof Electro-anatomic mapping systems in arrhythmias Europace, November 1, 2008; 10(suppl_3): iii28 - iii34. [Abstract] [Full Text] [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2008 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |